To calibrate 0% Saturation:

1. Rinse the probe well with DI rinse water. For best

accuracy, wipe the end of the probe dry. Do not touch the

membrane.

2.

Press the MODE key to select the % saturation mode.

3.

Place the probe in zero oxygen solution. Stir the solution

gently using the probe to homogenize the sample. Make

sure that the sample is continuously flowing past the

membrane sensor.

4.

Press the CAL key. The CAL indicator will appear above

the primary display. The primary display will show the

factory default % saturation measurement and the

secondary display will show 0.0.

5.

Wait for the reading to stabilise. If the READY indicator

feature is enabled, it will appear when the reading is

stable. (See section 8.7.1 on page 29 for Ready indicator

setup).

6.

Press the ENTER key. The meter automatically calibrates

to 0.0% saturation and returns to Measurement mode.

See figure on right.

NOTE

The factory default % saturation measurement in step 4 must read at 10% or under for the

calibration to work correctly. Whenever an error occurs during calibration, the ERR indicator

appears in the lower left hand corner of the display.

Figure 7: To calibrate for

0.0% saturation