16 FACTORY DEFAULT SETTINGS

Resetting the meter to factory default settings clears all DO calibration data and memory, and

returns other setup functions to the default settings shown in the table below. However, the

clock settings will remain as you have set them. See page 35 for directions on setting the

meter to factory default settings.

Program Function

Options/settings

Default

CLr

Memory clear

Yes / no

No

dPr

Pressure inputs

Hg or Pa

Hg

Barometric pressure

Adjust from 500 – 1499 mm
Hg

760 mm Hg

Salinity adjustment

Adjust from 0 – 50.00 ppt

0 ppt

OFS

Set % saturation offset

Up to ±10%

No offset

CAL

View calibration data

----

----

ELE View

probe

slope

----

1.000

View % saturation offset

----

0% offset

View mV = 100% saturation

----

37.0 mV

View mV = 0% saturation

----

0.3 mV

COF

Ready indicator / auto endpoint

On / off

Ready ON only

Select mg/L or ppm units

mg/L or ppm

mg/L

Select temperature units

°C or °F

Retains settings

LCd

Adjust back lit display

Levels 0 to 8 (brightest)

Retains settings

CLO

Setting clock

Year, date, time

Retains settings

rSt

Reset to factory default

Yes / no

no

NOTES:

The P7.0 clock function retains the year and time that you set, even after reverting to factory

default settings.