16 factory default settings – Thermo Fisher Scientific CyberScan DO 300 User Manual
Page 54

Instruction Manual
DO 300
51
16 FACTORY DEFAULT SETTINGS
Resetting the meter to factory default settings clears all DO calibration data and memory, and
returns other setup functions to the default settings shown in the table below. However, the
clock settings will remain as you have set them. See page 35 for directions on setting the
meter to factory default settings.
Program Function
Options/settings
Default
CLr
Memory clear
Yes / no
No
dPr
Pressure inputs
Hg or Pa
Hg
Barometric pressure
Adjust from 500 – 1499 mm
Hg
760 mm Hg
Salinity adjustment
Adjust from 0 – 50.00 ppt
0 ppt
OFS
Set % saturation offset
Up to ±10%
No offset
CAL
View calibration data
----
----
ELE View
probe
slope
----
1.000
View % saturation offset
----
0% offset
View mV = 100% saturation
----
37.0 mV
View mV = 0% saturation
----
0.3 mV
COF
Ready indicator / auto endpoint
On / off
Ready ON only
Select mg/L or ppm units
mg/L or ppm
mg/L
Select temperature units
°C or °F
Retains settings
LCd
Adjust back lit display
Levels 0 to 8 (brightest)
Retains settings
CLO
Setting clock
Year, date, time
Retains settings
rSt
Reset to factory default
Yes / no
no
NOTES:
The P7.0 clock function retains the year and time that you set, even after reverting to factory
default settings.