4.2 Temperature

Calibration

DO in mg/L is dependent on temperature, so it is first necessary

to verify the temperature reading. Temperature calibration is

accessible from the mg/L or ppm measurement mode.

The built-in temperature sensor in the probe is factory calibrated.

Calibrate your sensor only if you suspect temperature errors may

have occurred over a long period of time or if you have a

replacement probe.

1. Dip the DO probe into a solution with known temperature.

Allow sufficient time for the temperature to stabilise.

NOTE: Use an accurate temperature meter or bath to determine

the solution’s temperature.

2. Ensure that the meter is in measurement mode. Press MODE

key if necessary to select mg/L mode.

3. Press CAL key to go into calibration mode of mg/L. The

primary display shows the measured DO in mg/L and the

secondary display shows the temperature.

4. Press MODE key to go into temperature offset adjustment

mode. The primary and secondary displays will show the

temperature values.

5. Press MI/

or MR/

key to set the correct temperature value.

NOTE: Temperature calibration is restricted to ± 5°C from the

factory default value displayed during calibration (shown in the

secondary display).

6. Press ENTER key to confirm the adjusted temperature.

NOTE: To exit from Temperature Calibration mode without

confirming calibration, DO NOT press ENTER in Step 6. Press

CAL/MEAS instead.

Since temperature readings affect the accuracy of the mg/L

measurements, it is strongly recommended to carry out a mg/L

calibration after a temperature calibration is done.

(The values shown above are for illustration purpose. Your meter may display different values depending on the settings &

your environmental conditions)