Instruction Manual

DO 300

29

8.7.1

READY Indicator and Auto Endpoint Function

The first program lets you select:

“READY indicator ON” to indicate when the

reading is stable.

“READY indicator OFF” for faster meter response.

Auto Endpoint Function ON. Select auto

endpoint ON to “HOLD” the reading when it is

stable for more than 5 seconds. The display

automatically freezes, and the HOLD indicator

appears on the left side of the display. Press the

HOLD key to release the display and access other

functions.

From measurement mode

1.

Press SETUP key to enter SETUP mode.

2. Press

MI/

or MR/

keys to scroll through main

groups until you view parameter COF in the upper

display.

3. Press the ENTER key to select parameter “rdY”

(Ready).

4. Press MI/

or MR/

keys to select the

configuration you require.

ON switches the READY indicator on.

OFF switches the READY indicator off.

• ON and HOLD together switches the auto

endpoint feature on.

5.

Press the ENTER key to confirm selection and to

proceed to:

In % saturation SETUP mode: step 3 on Section 8.7.3.

In mg/L (ppm) SETUP mode: step 4 on Section 8.7.2.

Or press the CAL/MEAS key twice to return to measurement mode.

NOTES: Meter default is set for Ready indicator ON, and auto endpoint function off.

Figure 27: To configure for
READY and Auto Endpoint

functions