Thermo Fisher Scientific CyberScan DO 300 User Manual
Page 32

Instruction Manual
DO 300
29
8.7.1
READY Indicator and Auto Endpoint Function
The first program lets you select:
•
“READY indicator ON” to indicate when the
reading is stable.
•
“READY indicator OFF” for faster meter response.
•
Auto Endpoint Function ON. Select auto
endpoint ON to “HOLD” the reading when it is
stable for more than 5 seconds. The display
automatically freezes, and the HOLD indicator
appears on the left side of the display. Press the
HOLD key to release the display and access other
functions.
From measurement mode
1.
Press SETUP key to enter SETUP mode.
2. Press
MI/
or MR/
keys to scroll through main
groups until you view parameter COF in the upper
display.
3. Press the ENTER key to select parameter “rdY”
(Ready).
4. Press MI/
or MR/
keys to select the
configuration you require.
•
ON switches the READY indicator on.
•
OFF switches the READY indicator off.
• ON and HOLD together switches the auto
endpoint feature on.
5.
Press the ENTER key to confirm selection and to
proceed to:
•
In % saturation SETUP mode: step 3 on Section 8.7.3.
•
In mg/L (ppm) SETUP mode: step 4 on Section 8.7.2.
Or press the CAL/MEAS key twice to return to measurement mode.
NOTES: Meter default is set for Ready indicator ON, and auto endpoint function off.
Figure 27: To configure for
READY and Auto Endpoint
functions