5 MEASUREMENT

5.1

Taking DO readings

During measurement, the probe can be:

z

Fully immersed in the solution

z

Partially immersed in the solution

Do not allow the probe’s membrane surface to touch anything!

It is always advisable to insert the probe guard (refer to Figure

34 and 35 on pages 38 and 39)

IMPORTANT: Since the DO probe consumes oxygen from the

sample, the sample must constantly flow past the membrane

to achieve more accurate readings. You can use a stirrer to

accomplish this.

To take measurements:

1.

Rinse the probe well with DI rinse water or rinse solution.

2. Select the appropriate measurement mode. Press the

MODE key to toggle between modes:

z

% Saturation

z

mg/L (or ppm)

z

date/time clock

3.

Dip the probe into the sample. Stir the solution gently

using the probe to homogenise the sample. Make sure

that the sample is continuously flowing past the

membrane sensor.

4. Note the reading on the display. If the READY

indicator is switched on, it will appear when the reading

is stable.

5.

You can press the

(LIGHT) key to activate the backlit

LCD. This helps you read measurements in dimly lit areas.

Figure 9: Change

measurement mode

Figure 10: Taking

measurement