beautypg.com

Thermo Fisher Scientific CyberScan DO 300 User Manual

Page 13

background image

Instruction Manual

DO 300

10

4.4

Dissolved Oxygen Calibration in mg/L (ppm) mode

Calibrating the meter to 100% in saturation mode will also calibrate the mg/L mode at the

value in mg/L corresponding to 100% saturation. This should produce acceptable results in

most applications.

This meter also lets you make a calibration adjustment in mg/L (ppm) mode without affecting

your % saturation calibration.

To select between mg/L and ppm units, see page 30.

To calibrate in mg/L (ppm) mode:

1. Calibrate 100% saturation as per Section 4.3.

2. Rinse the probe well with DI rinse water or rinse solution.

For best accuracy, wipe the end of the probe dry. Do not

touch the membrane.

3. Press the MODE key to select the mg/L (ppm) mode.

4. Dip the probe into a sample of known oxygen concentration.

Wait for the reading to stabilise. If the READY indicator

feature is enabled, it will appear when the reading is stable.

5. Press the CAL key. The CAL indicator will appear above the

primary display. The primary display will show the factory

default mg/L (ppm) measurement and the secondary

display shows the temperature.

6. Press the MI/

or MR/

keys to set the reading according

to the sample value.

7. Press the ENTER key. The meter automatically calibrates to

the value you entered and returns to Measurement mode.

Notes on mg/L (ppm) mode calibration

• During mg/L (ppm) measurement and calibration, the

meter adjusts to the barometric pressure value and the

salinity value that is programmed into the meter

• You can change the barometric pressure value and

salinity value in the mg/L (ppm) SETUP mode (see

section 8.3 on page 21 for directions).

Whenever an error occurs during calibration, the ERR

indicator appears in the lower left hand corner of the

display.

Figure 8: To calibrate in

mg/L (ppm) mode

(The values shown above

are for illustration purpose.

Your meter may display

different values depending

on the settings & your

environmental conditions)