Instruction Manual

DO 300

45

11 ERROR

MESSAGES

LCD Display

Indicates

Cause

Solution

Err annunciator

Unrecognised input
from keypad

Wrong input in selected
mode.

Release key. Select valid
operations depending on
mode.

CAL & Err
annunciators on /
Buffer and
electrode
indicators blink.

Calibration error.

Wrong value input at
calibration.

Dirty probe.

Check your input value,
clean probe.

See Calibration sections or
Probe Maintenance
section.

Battery indicator
blinks

Low battery level.

Need new batteries or
battery connection is
bad.

Clean battery contacts.
Replace batteries with fresh
ones, noting polarity.

If an error message appears in the primary display (the upper row of larger digits), switching off the

meter and switching it on again may eliminate the error message.

If error persists, or the meter shows incorrect values, return the meter. See Warranty and Return of

Items sections.

For a complete diagram of the display, see page 2.