ELE: Electrode Properties

These “view only” parameters show you the electrode properties

for diagnostic purposes:

A.

Probe Slope: Lets you view and give an indication of the

probe’s efficiency. The value displayed is the ratio of the

theoretical value to the actual value produced by the probe.

The higher the number, the lesser output from the probe. The

ratio displays from 0.5 to 1.999.

B.

% Saturation Offset: (available in % saturation SETUP mode

only): Lets you see the value of the % saturation offset entered

in parameter “OFS” (see page 24 for instructions).

C.

100% Saturation mV value: Lets you view the sensor’s mV

output corresponding to 100% saturation.

D.

0% Saturation mV Value: Lets you view the sensor’s millivolt

output corresponding to 0% saturation.

From measurement mode:

1.

Press the MODE key to select the measurement mode for the

electrode properties you want to view: % or mg/L (ppm).

2.

Press the SETUP key to enter SETUP mode.

3. Press

MI/

or MR/

keys to scroll through main groups until

you view parameter “ELE” in the upper display.

4.

Press the ENTER key. The upper display shows the probe

slope.

5. Press the ENTER key. The upper display shows the %

saturation offset. See directions for setting this offset on page

24. NOTE: This % saturation slope and offset appears only in

% saturation SETUP mode. If you are in mg/L (ppm) SETUP

mode, the meter skips step 4 and 5 to show step 6 on page 27.

Figure 23: View electrode

properties from %

saturation SETUP mode

Checking electrode data
from % saturation
SETUP mode

SETUP

p 5.0

ele

SETUP

.948

el 1

%

K=

SETUP

EL 2

7.6

%

SETUP

EL 3

39.2

mV

ENT

ER

EN

TER

EN

TER

SETUP

EL 4

- 0.1

mV

EN

TER