Thermo Fisher Scientific CyberScan DO 300 User Manual

Page 26

background image

Instruction Manual

DO 300

23

8.3.2

Salinity Adjustment Mode

NOTE: This mode appears only from mg/L (ppm)

SETUP mode .

Salinity correction mode lets you correct for the

variations in oxygen solubility due to salt concentration

in the sample.

From measurement mode:

1.

Press the MODE key to select mg/L (ppm) mode.

2.

Press the SETUP key to enter SETUP mode.

3. Press

MI/

or MR/

keys to scroll through main

groups until you view parameter “dPr” in the upper

display.

4.

Press the ENTER key five times. The upper display

shows the salinity value and the lower shows SAL.

5.

Press

MI/

or MR/

keys to enter the correct

salinity adjustment factor. The salinity adjustment

factor range is 0.0 to 50.0 ppt.

6.

Press the ENTER key to confirm selection and to

move back to main group “dPr”. If you want to

return to measurement mode, press CAL/MEAS.

(The values shown above are for illustration purpose only)

Figure 19: Input Salinity values for

automatic correction