Techne 3PrimeG User Manual
Page 38

37
Ramp rate
• Touch the ramp rate setting to adjust the ramp
rate.
The range is 0.1ºC/s (minimum) to 3.0ºC/s
maximum for heating and 2.1ºC/s for cooling. The
value can be entered to 1 decimal place. Ramp
rate will affect the heating or cooling rate from the
previous step to the current step.
• Once set, touch OK to return to the
Programming Screen.
Delete a stage
• To delete a stage, touch the stage area to
highlight it and then touch Delete.
INSERT/DELETE A STEP
Insert a new step
• Touch Insert a new step to insert the next step
in the current stage.
• Edit the new step temperature, hold time and
ramp rate as required.
Delete a step
• To delete a step, touch the area near to the
step number to highlight it and then touch
Delete. More than one step can be selected at
a time.
INSERT/DELETE A PAUSE
Insert a pause
• To insert a pause between stages, touch Insert
a pause.
The hold temperature will be that of the previous
step. The unit will not continue with the remainder
of the program until it is manually un-paused by the
user.
Delete a pause
• To delete a pause, touch the pause button to
highlight it and then touch Delete.
Note: To insert a pause between two previously
programmed stages, it is necessary to first delete
the later stages to where the pause is required.
Continue adding stages and steps as required. Use the up and down arrow buttons to scroll through
the program as necessary.