Techne 3PrimeG User Manual
Page 34

33
Pause
This will cause the unit to pause before the program
starts at the ambient temperature of the block (unless
Sample Cooling has been set to On when the unit will
pause with the block held at 4ºC). The program will not
commence until the unit is manually un-paused.
• Touch On/Off to toggle between settings.
Initial Denaturation
Use this to denature the sample and/or activate the
enzyme before the thermal cycling program begins.
• Touch On/Off to toggle between settings.
• Touch the time button to set the required hold time.
• Touch the temperature button to set the required
temperature.
Hot Start
This will cause the unit to pause at the set temperature.
The program will not continue until the unit is manually
un-paused.
• Touch On/Off to toggle between settings.
• Touch the temperature button to set the required
temperature.
• Touch More Settings for further program defaults or
Back to return to the Settings menu.
Final Extension
Use this to add a final extension step to the program to
ensure complete amplification of the products.
• Touch On/Off to toggle between settings.
• Touch the time button to set the required hold time.
• Touch the temperature button to set the required
temperature.
Final Hold
This will hold the block at the set temperature at the
end of a program until the user manually stops the unit.
• Touch On/Off to toggle between settings.
• Touch the temperature button to set the required
temperature.
• Touch More Settings to return to the top of the
Program Defaults menu or Back to return to the
Settings menu.