18

6. With the

3

PrimeX and G, the heated lid has a

rotating knob to adjust the lid pressure on the

samples, allowing for a variety of consumables to

be used. To adjust the pressure:

a. Place the samples in the block and place the lid

in the down position.

b. Adjust the orange rotating knob anticlockwise

to give the least pressure then close the lid

fully.

c. Gently rotate the knob clockwise until you can

feel pressure beginning to be applied.

d. Rotate the knob a further quarter of a turn and

this will give the correct pressure. Do not over-

tighten.

7. Once the pressure has been set it does not need to be adjusted unless a different block or

consumable is used.