SHIMPO DAG-2000 User Manual
Page 9
DIGITAL AIR MICROMETER USERS MANUAL (Second Edition)
9
3.3 For first time measurement tool connection cases
Shown here is the procedure up to measurement wherein measurement tool is connected to the
main body for the first time.
Note: There are cases wherein detector adjustment is not needed depending on the measurement
tool.
Turn ON compressed air
Turn ON power
Switch to settings mode
Change the settings
Switch to master
calibration mode
(detector adjustment)
Detector adjustment
Switch to master
calibration mode (master
calibration)
Master calibration
Measurement
•
Switch from measurement mode to settings mode
Because of work tolerance change, change the judgment
limit values.
Settings value change can be done in setting mode.
Please refer [Chapter 6, 6.4 Settings] for details
•
At settings mode, input the settings value for the
measurement tool that will be connected.
Please refer to [Chapter 6, 6.4 Settings] for details
•
Switch to measurement mode after finishing settings
change
Please refer to [Chapter 6, 6.4 Settings] for details
•
Switch from measurement mode to detector adjustment
Please refer to [Chapter 6, 6.5 Detector Adjustment] for
details
•
Switches from detector adjustment to master calibration
Refer to [Chapter 6, 6.5 Detector Adjustment] for
details.
•
Adjustment of detector through ZERO and sensitivity
tuner.
Please refer to [Chapter 6, 6.5 Detector Adjustment] for
details