SHIMPO DAG-2000 User Manual
Page 12
DIGITAL AIR MICROMETER USERS MANUAL (Second Edition)
12
Large range master calibration
(Sensitivity correction)
Master calibration completion
⑥ After “MAX M.” shown in the English display for
2 seconds, data after ZERO correction will be shown.
Please set the large range master to the measurement tool.
ZERO correction data will be shown in the main display.
ZERO correction data display colors are:
• Green - sensitivity master calibration is possible.
Please press ENT. Sensitivity correction will be
performed.
• Red - sensitivity master calibration is not possible.
Detector adjustment will be performed.
Please press /
MAS
.
“ADJ.” will be shown in the English display and will
switch to detector adjustment.
Please refer to [Chapter 6, 6.5 Detector Adjustment] for
details.
“MAS OK” will be shown in the English display and will
switch to measurement mode.
Master calibration has error if anything else is displayed.
Please press ENT and repeat from .
About the master calibration error, please refer to [3.4 Master
calibration Error].