SHIMPO DAG-2000 User Manual
Page 19
DIGITAL AIR MICROMETER USERS MANUAL (Second Edition)
19
(3) Settings details
This explains the settings item name and settings details at the settings mode.
Settings mode is mainly divided into the following 3:
• Normal
[Standard settings] --- master values, judgment limit values
can be set
• Detail
[Detailed settings] --- measurement range, display analysis function, polarity
ditto
• System
[System settings] --- display data, moving average
ditto
① Settings item names and settings details at Normal [Standard settings]
Settings concerning master values
• [MIN M.] --- Please input the small range master values used in master calibration (2 masters).
• [MAX M.] --- Please input the large range master values used in master calibration (2 masters).
• [ZERO M.] --- Please input the ZERO master values used in master calibration (1 master).
Settings concerning judgment limit values
• [-NG/-OK] --- Please input the limit values of –NG and -OK
• [-OK/ OK] --- Please input the limit values of –OK and OK
• [ OK/+OK] --- Please input the limit values of OK and +OK
• [+OK/+NG] --- Please input the limit values of +OK and +NG
In case judgment of –OK is not needed, please set the same settings values to [-NG/-OK] and
[-OK/OK] and in case judgment of +OK is not needed, please set the same settings values to
[OK/+OK] and [+OK/+NG].
Settings concerning master
• [CORRECT] --- Machine difference correction values
can be set
Adds and displays the above data to the master correction data.
② Settings item name and settings details at Detail [Detailed settings]
• [RANGE] --- Measurement range
can be selected
10
µm is optional.
• [RESOLUTION] --- display analysis function
ditto
Settings concerning detector
• [POL] --- Polarity
Please select + for inner diameter measurement and – for outer dimension measurement
• [GAIN] --- Sensitivity rough adjustment values
can be set
Normally, settings values is fixed depending on the selected measurement range.
Please input:
24 if measurement range 100
µm is selected
33 if measurement range 50
µm is selected
68 if measurement range 20
µm is selected
(204 if measurement range 10
µm is selected)
In case measurement range is changed, the above values will be set.
• [CONSTANT] --- Sensitivity fine adjustment constant can be set
Normally, please input 1.000.
Settings concerning master
• [CalMode] --- Master calibration method
can be selected
Please select:
[MIN & MAX] for master calibration by small range or large range master.
[ZERO M.] for master calibration by ZERO master.
③ Settings item name and settings details at System [System settings]
• [MEAS SW] --- External button input movement
can be selected
• [SMOOTH] --- Moving average
ditto
Settings concerning display
Data shown in:
• [MainDisp] --- Main display
can be selected
• [CharDisp] --- English display
ditto