DIGITAL AIR MICROMETER USERS MANUAL (Second Edition)

5

1.3 Block Diagram

(1) Structure

Arrow represents the direction of air and electric signal.

is optional.

(2) Operation flow

Shown below is the software operations diagram starting from turning on power.

Compressed air

purifier

Power ON

Initialization mode

Measurement mode

Program switching mode

Setting mode

Master calibration mode

• Measurement range selection

• Input of master value
• Input judgment limits

• Master calibration

• Adjustment of detector

• Master calibration data clear

Program switching
Can save up to 10 kinds of
settings values

• Reading of last setting values used

and master calibration data

• Confirmation of indicator light

(visual)

A/E Converter

CPU

Memory for saving settings
values and master calibration
data (good for 10 types)

Optional

DC I/O base

Memory for
operation software

Industrial pressure

Air circuit

Pressure sensor

Amp circuit

Parallel I/O

Work and master

Measurement
element/tool

Input

Range switch

RS232C

Digimatic output

External input button

In case industrial
pressure contains
moisture and oil

Regulator

0.196

±0.005MPa

0.3~0.7MPa

Mist separator

Filter