Measurement Modes

Peak, Mean Measurement modes are available on the TRC:

To select the measuring mode required for testing, follow the settings below.

5.1

Peak Mode

The Peak value displayed on the TRC is measured at 1,000 times per second. (1000Hz)

The Peak value is reset by pressing the Reset key or activating the Auto reset function (refer to 4.1.3

Auto Reset Setting).

Set zero by pressing Zero key.

5.2

Mean Mode

The mean mode is the real time value based on the average data captured in 1,000 samples per second.

(1000Hz)

F03

Display Update time

Average

1

1 time/sec

Displays the average of 1,000 measuring data by 1msec sampling

2

2 times/sec

Displays the average of 500 measuring data by 1msec sampling

4

4 times/sec

Displays the average of 250 measuring data by 1msec sampling

8

8 times/sec

Displays the average of 125 measuring data by 1msec sampling

Mean Mode

Mean

Peak

Peak Mode

Mean

Peak

lbI

lbI

Mode

Mode

Mean

Peak

Press and hold Zero key

until the display shows

“0”

Mean

Peak

lbI

lbI

Zero