9

Technical data

9.0.1

Input

Measured variable

Temperature (temperature linear transmission behaviour), resistance and voltage

Measuring range

The transmitter records different measuring ranges depending on the sensor connection and input
signals.

Input

Designation

Measuring range limits

Min. span

Resistance thermometer (RTD)
To IEC 60751
(

α = 0.00385)

To JIS C1604-81
(

α = 0.003916)

To DIN 43760
(

α = 0.006180)

To Edison Copper Winding No.15
(

α = 0.004274)

To SAMA
(

α = 0.003923)

To Edison Curve
(

α = 0.006720)

To GOST
(

α = 0.003911)

To GOST
(

α = 0.004278)

Pt100
Pt200
Pt500

Pt1000

Pt100

Ni100

Ni1000

Cu10

Pt100

Ni120

Pt50

Pt100

Cu50, Cu100

Polynomial RTD

Pt100 (Callendar - van Dusen)

-200 to 850 °C (-328 to 1562 °F)
-200 to 850 °C (-328 to 1562 °F)
-200 to 250 °C (-328 to 482 °F)
-200 to 250 °C (-238 to 482 °F)

-200 to 649 °C (-328 to 1200 °F)

-60 to 250 °C (-76 to 482 °F)
-60 to 150 °C (-76 to 302 °F)

-100 to 260 °C (-148 to 500 °F)

-100 to 700 °C (-148 to 1292 °F)

-70 to 270 °C (-94 to 518 °F)

-200 to 1100 °C (-328 to 2012 °F)

-200 to 850 °C (-328 to 1562 °F)

-200 to 200 °C (-328 to 392 °F)

-200 to 850 °C (-328 to 1562 °F)
-200 to 850 °C (-328 to 1562 °F)

10 K (18 °F)
10 K (18 °F)
10 K (18 °F)
10 K (18 °F)

10 K (18 °F)

10 K (18 °F)
10 K (18 °F)

10 K (18 °F)

10 K (18 °F)

10 K (18 °F)

10 K (18 °F)
10 K (18 °F)

10 K (18 °F)

10 K (18 °F)
10 K (18 °F)

• Type of connection: 2-wire, 3-wire or 4-wire connection
• With 2-wire circuit, compensation of wire resistance possible (0 to 30

Ω)

• With 3-wire and 4-wire connection, sensor wire resistance to max. 50

Ω per wire

• Sensor current:

≤ 0.3 mA

Resistance transmitter

Resistance

Ω

10 to 400

Ω

10 to 2000

Ω

10

Ω

100

Ω

Thermocouples (TC)
To NIST monograph 175,
IEC 584

to ASTM E988

to DIN 43710

Type B (PtRh30-PtRh6)

1)

Type E (NiCr-CuNi)
Type J (Fe-CuNi)
Type K (NiCr-Ni)
Type N (NiCrSi-NiSi)
Type R (PtRh13-Pt)
Type S (PtRh10-Pt)
Type T (Cu-CuNi)

Type C (W5Re-W26Re)
Type D (W3Re-W25Re)

Type L (Fe-CuNi)
Type U (Cu-CuNi)

0 to +1820 °C

(32 to 3308 °F)

-270 to +1000 °C (-454 to 1832 °F)
-210 to +1200 °C (-346 to 2192 °F)
-270 to +1372 °C (-454 to 2501 °F)
-270 to +1300 °C (-454 to 2372 °F)

-50 to +1768 °C (-58 to 3214 °F)
-50 to +1768 °C (-58 to 3214 °F)

-270 to +400 °C (-454 to 752 °F)

0 to +2320 °C (32 to 4208 °F)
0 to +2495 °C (32 to 4523 °F)

-200 to +900 °C (-328 to 1652 °F)
-200 to +600 °C (-328 to 1112 °F)

500 K (900 °F)

50 K (90 °F)
50 K (90 °F)
50 K (90 °F)
50 K (90 °F)

500 K (900 °F)
500 K (900 °F)

50 K (90 °F)

500 K (900 °F)
500 K (900 °F)

50 K (90 °F)
50 K (90 °F)

• Internal cold junction (Pt100); accuracy of cold junction: ± 1 K
• Max. sensor resistance 10 k

Ω (if sensor resistance is greater than 10 kΩ, error message as per NAMUR NE 89)