Maintenance
Series 642
28
Pyromation, Inc.
7
Maintenance
No special maintenance work is required on the device.
8
Trouble-shooting
8.1
Trouble-shooting instructions
Always begin trouble-shooting with the following checklists if errors occur after commissioning or
during measuring operation. The questions guide you to the cause of the error and the appropriate
remedial action.
8.2
Error messages
Fault code
Cause
Action/cure
Mode
1)
0
No error, warning
-
-
10
Hardware error (device defective)
Replace device
F
13
Reference measuring point defective
Replace device
F
15
EEprom defective
Replace device
F
16
A/D converter defective
Replace device
F
17
Ambient temperature limit overshot
Electronics possibly damaged by
overshooting ambient temperature limits,
return electronics to manufacturer for
checking
0, F
19
Supply voltage too low
Check supply voltage; check connection
wires for corrosion
F
50
Sensor cable open circuit
Check sensor
*
51
Sensor short-circuit
Check sensor
*
52
Sensor corrosion
Check sensor
*
53
Outside sensor range
Wrong sensor type for application
*
81
Alarm: measuring range undershoot
Measuring range poss. set too small
F
82
Alarm: measuring range overshoot
Measuring range poss. set too small
F
106
Warning: upload/download active
-
C
107
Warning:
Output simulation active
Deactivate output simulation
C
201
Warning:
Measured value too low
Change PV lower range value
M
202
Warning:
measured value too large
Change PV upper range value
M
203
Warning:
Ambient temperature limit overshot
Electronics possibly damaged by
overshooting ambient temperature limits,
return electronics to manufacturer for
checking
0
206
Warning:
Sensor corrosion
Check sensor
M
208
Reset device to factory default values
-
0