Pyromation Series 642 User Manual
Page 25
Series 642
Commissioning
Pyromation, Inc.
25
Function group SERVICE FUNCTIONS
Availability in TransComm Light, HART
®
communicator DXR375 (symbol
7
)
TransComm Light
7
Security locking
Set-up release code.
Input:
• Lock = 0
• Release = 246
+
+
Reset to default
Reset to factory default values.
Input: 642
0
+
+
Output simulation
Activate simulation mode.
Input:
• OFF
• ON
+
+
Simulation value
Input of the simulation value (current).
Input: 3.58 to 23 mA
+
+
User calibration
(trim) analog output
For changing the 4 or 20 mA value by ± 0.150 mA
• Trimming 4 mA
• Trimming 20 mA
+
+
Function group MEASURED VALUES
Availability in TransComm Light, HART
®
communicator DXR375 (symbol
7
)
TransComm Light
7
PV
PV value
+
+
AO
PV value in mA
-
+
PV %
PV value in %
-
+
Sensor
Sensor process value
-
+
Internal temperature
Internal temperature of the device
-
+