Martel Electronics DMC1400 User Manual
Page 82
1)
The test outputs can not be changed, they will be the predefined
values that were downloaded. For pressure and manual output
types, the downloaded value is displayed for the technician to
set it on the external source. For all other output types, the
calibrator automatically generates the As Found value.
2)
Once the SAVE function key is pressed for the last test point, the
process automatically proceeds to the identification data entry
stage.
3)
A REPEAT function key is available for restarting the test at the
first point.
8.4.3 Identification Data Entry
At this step the technician is given the opportunity to update any of the
downloaded tag identification fields, except for the tag number which
must remain the same in order for the PC program to synchronize the
test results with its database during upload. Otherwise this process is
identical to that described under New Tag As Found identification data
entry. When the SAVE function key is pressed, the process
automatically proceeds to the test conclusion stage.
8.4.4 Test Conclusion
If test validation was selected, a display showing the overall pass/fail
status of the test is displayed with a prompt asking whether details
should be displayed. Press the YES function key to display the
details, or the NO function key to skip the details. In either case the
adjustment prompt is displayed next.
A prompt is displayed asking whether further adjustment and testing
is required. Press the YES function key to proceed to the adjustment
menu, or the NO function key to proceed to the first level document
mode menu.
8.5 As Left Test
An As Left test can immediately follow an As Found test or a previous
As Left test, or it can be initiated at a later time for any tag stored in
memory. If it immediately follows another test, no tag selection step
is required and data collection starts right away. If it is initiated at a
later time, the tag must be selected first.
8.5.1 Tag Selection
At the main menu, press the MENU function key, followed by the
MORE function key, followed by the DOCUMENT function key to
display the first level document mode menu.
80