beautypg.com

Martel Electronics DMC1400 User Manual

Page 79

background image

If pressure is selected on the upper or lower display, the pressure
zero prompt is displayed first and operates as described under As
Found Setup.

The prompt displayed during the test depends on the input and
output types selected. Any combination which includes pulse train
output has its own unique prompt sequence. All other combinations
are covered by a generic prompt.

Generic Type Prompt

A prompt similar to the following is displayed for each test point
when the output is not pulse train.

The PT in the upper right corner of the display indicates that this is
the adjustment display. If an As Found data point is being
generated, its number is displayed. If a user entered value is being
generated, two dashes are displayed.

Press the UP or DOWN function key to cycle through the As Found
test points. If the output type is not pressure or manual, the
calibrator will generate it. If the output type is pressure or manual, the
output value is displayed so that it can be set on the external source.

If the output type is not pressure or manual, any value different from
the test points can be entered using the numeric keys, followed by
ENTER.

Press the AS LEFT function key to proceed directly to As Left data
collection.

Pulse Train Type Prompt

When the output is pulse train, the following is displayed for each
test point.

77