8.2.3 Identification Data Entry
When the DONE function key is pressed at the data collection stage,
the first page of the following two page identification data display is
shown. Initially all data fields are blank, but after data is entered the
current value of each field is shown to the right of its description.
These two pages provide space to enter eight phrases of up to 16
characters each to describe the test that has been performed. The
first seven phrases are predefined to be Tag Number, Instrument
Manufacturer, Model Number, Instrument Serial Number, Technician
Name or ID number, Temperature, and Humidity. The last phase,
Other is for client use. There is no restriction on the data that may be
entered except for Tag Number which must have a non blank first
character and must be unique among all tags currently stored in the
calibrator.
Press the NEXT function key to switch between the two pages.
Press the SAVE function key to save the data and proceed to the
results validation step.
Press the UP ARROW and DOWN ARROW keys to move the field
selector highlight from one field to the next. To enter data into the
highlighted field, press the SELECT function key and a prompt
similar to the following is displayed showing the field name and
current value on the top line.