Specifications, Table 10: dc voltage measurement/source, Table 11: dc ma measurement/source – Martel Electronics DMC1400 User Manual
Page 100: Table 12: frequency measurement/source

10. Specifications
All measurements apply at 23°C ± 5°C. unless specified otherwise.
Outside of this range the stability of the measurements is ±
0.005%of reading/°C.
Table 9: General Specifications
Operating Temperature
-10°C to 50°
Storage Temperature
-20°C to 70°C
Power
4 X AA batteries; Alkaline or optional rechargeable
Low battery warning
Yes
Serial Communications
Yes, ASCII
CE - EMC
EN50082-1: 1992 and EN55022: 1994 Class B
Safety
CSA C22.2 No. 1010.1: 1992
Table 10: DC Voltage Measurement/Source
Range
Accuracy
(% of reading ± floor)
Read: Isolated(Upper Display)
0.000V- 30.000V
0.015% ± 2mV
Read: non-Isolated(Lower Display)
0.000V - 20.000V
0.015% ± 2mV
Source
0.000V - 20.000V
0.015% ± 2mV
Maximum current output in voltage ranges is 3mA with an output impedance of <= 1
W.
Table 11: DC mA Measurement/Source
Range
Accuracy(% of reading
± floor)
Read: Isolated(Upper Display)
0.000mA - 24.000mA
0.015% ± 2
mA
Read: non-Isolated(Lower Display)
0.000mA - 24.000mA
0.015% ± 2
mA
Source
0.000mA - 24.000mA
0.015% ± 2
mA
Maximum load on mA source is 1000
W. Voltage input range on simulate mode 5V - 30V.
Table 12: Frequency Measurement/Source
Range
Accuracy
(% of reading ± floor)
Read
2.0CPM - 600.0CPM
0.05% ± 0.1CPM
1.0Hz - 1000.0Hz
0.05% ± 0.1Hz
1.00KHz - 10.00KHz
0.05% ± 0.01KHz
Source
2.0CPM - 600.0CPM
0.05%
1.0Hz - 1000.0Hz
0.05%
1.00KHz - 10.00KHz
0.125%
Input voltage amplitude range on frequency is 1V to 20V zero based square wave only.Output amplitude
is adjustable from 1V to 20V, and is a square wave with 50% duty cycle.For output frequency, a slight
negative offset of approximately -0.1V is present to assure zero crossing.
98