Martel Electronics DMC1400 User Manual
Page 37
35
Event Status Enable Register (ESE)
Enables and disables bits in the ESR. Setting a bit to 1 enables the
corresponding bit in the ESR, and setting it to 0 disables the
corresponding bit. Cleared at power reset. Bit assignments for the
ESR and the ESE respectively are shown below.
15
14
13
12
11
10
9
8
0
0
0
0
0
0
0
0
7
6
5
4
3
2
1
0
PON
0
CME
EXE
DDE
QYE
0
OPC
PON
Power On. Set to 1 if power was turned on and off before
the Event Status Register was read.
CME
Command Error. Set to 1 when the calibrator receives an
invalid command. Entering an unsupported RTD type may
cause such an error.
EXE
Execution Error. Set to 1 when the calibrator runs into an
error while executing is last command. A parameter that
has too significant figures may cause such an error.
DDE
Device-dependent Error. Set to 1 when, for example, the
output of the calibrator is overloaded.
QYE
Query Error.
OPC
Operation Complete. Set to 1 when the calibrator has
finished executing all commands before the command
*OPC was entered.