Calibrator
Memory Operations
89
Recording Min and Max Measurements
You can set the display to record and show the
maximum and minimum readings. Min and Max
readings are always undamped, even if Dampen is
On. Press the
More Choices
softkey twice, then
press the
Min Max
softkey to turn on this feature.
Press the
c
key to reset the Min Max registers.
Press the
Min Max
softkey again to revert to the
normal display. The following figure shows the
display with Min Max on:
gj55s.eps
Running a Preloaded Task
Press the
More Choices
softkey until the
Tasks
softkey appear, then press
Tasks
to view the list of
tasks (procedures) downloaded from a host
computer. Tasks are calibrator configurations, saved
with a procedure name, for example the type and
manufacturer of a specific transmitter. A task
configures the calibrator for transmitter calibration
with all the calibration parameters (source and
measure functions, 0% and 100% levels, test
strategy) predefined.
While the task is controlling the calibrator, the
Continue
softkey becomes
Continue Task
.
Clearing Memory
In Setup mode, highlight the
Clear Memory
choice
and press
e
to clear all the memory:
•
Saved results
•
Min Max data
•
Log data sets
A confirmation message appears so that you do not