Martel Electronics MasterCAL User Manual
Page 93
Calibrator
Transmitter Mode
83
7.
Press
M
for SOURCE mode.
8.
Press the appropriate function key for the control
output (e.g.,
v
or
m
). If the transmitter is
connected to a current loop that has a power
supply, select
Simulate Transmitter
for the
current output choice.
9.
Select a source value, e.g., 4 mA.
10.
Press
M
for MEASURE/SOURCE mode.
11.
Press
More Choices
until the
Transmitter Mode
softkey appears.
12.
Press the
Transmitter Mode
softkey.
13.
Set the 0% and 100% values for MEASURE and
SOURCE on the display. You can select
Linear
or
√
for the transfer function.
14.
Press
Done
.
15.
The calibrator is now in Transmitter mode. It is
measuring the process input and sourcing the
control signal output proportional to the input.
16.
To change any of the Transmitter mode
parameters, press
Change Setup
, and repeat the
process in step 13.
17.
To exit Transmitter mode, press the
Abort
softkey.
- 3001 (94 pages)
- BetaGauge 301 (40 pages)
- BetaGauge 311A-Ex (32 pages)
- BetaGauge 311A (48 pages)
- BetaGauge 330 (48 pages)
- BetaGauge 3II (197 pages)
- BetaGauge PI (20 pages)
- BetaGauge PIR (24 pages)
- BetaLOG-PI (28 pages)
- BetaLOG-TI (16 pages)
- BetaProbe TI/TI+ (9 pages)
- DPC-30 (12 pages)
- DPC300A (48 pages)
- DMC1410 (108 pages)
- IVC-222HPII (12 pages)
- LC-110/H (50 pages)
- BETA LDT-500 (2 pages)
- M2000 (28 pages)
- MC1010 (46 pages)
- MC1210 (56 pages)
- TC100 (16 pages)
- T150 (8 pages)
- T140 Series (12 pages)
- PTC8010 (36 pages)
- PSC4010 (36 pages)
- MS-420 (1 page)
- MECP500 (2 pages)
- MECP2000 (2 pages)
- MECP10K (4 pages)
- MECP100 (2 pages)
- T130 (2 pages)
- T120 (2 pages)
- MC1200 (50 pages)
- MC1000 (28 pages)
- LC100 (20 pages)
- DMC1400 (106 pages)
- 320 (60 pages)
- 230 (106 pages)
- 130 (85 pages)
- QM100V (1 page)
- QM100V-40 (1 page)
- QM110T (2 pages)
- QM120ET (2 pages)
- QM130M (2 pages)