Measurement scale, Linear-output transmitters – Martel Electronics MasterCAL User Manual
Page 51
Calibrator
Using Measure Mode
41
Measurement Scale
This feature lets you scale the measurements in
accordance with a particular process instrument’s
response. Percent of scale works for linear-output
transmitters or square-law transmitters such as
differential pressure transmitters that report flow rate.
Linear-Output Transmitters
1.
If necessary, press
M
for MEASURE mode.
2.
Select a measurement function (
m
,
v
,
h
,
q
,
t
, or
p
) as previously described.
3.
Press the
Scale
softkey.
4.
Select
% scale
from the list.
5.
Use the numeric keypad to enter the 0% of scale
value (
0% Value
).
6.
Press
e
.
7.
Use the numeric keypad to enter the 100% of
scale value (
100% Value
).
8.
Press
e
.
9.
Press the
Done
softkey.
Percent of scale remains in effect until you change to
another measurement function or press the
Scale
softkey and select another scale mode.
- 3001 (94 pages)
- BetaGauge 301 (40 pages)
- BetaGauge 311A-Ex (32 pages)
- BetaGauge 311A (48 pages)
- BetaGauge 330 (48 pages)
- BetaGauge 3II (197 pages)
- BetaGauge PI (20 pages)
- BetaGauge PIR (24 pages)
- BetaLOG-PI (28 pages)
- BetaLOG-TI (16 pages)
- BetaProbe TI/TI+ (9 pages)
- DPC-30 (12 pages)
- DPC300A (48 pages)
- DMC1410 (108 pages)
- IVC-222HPII (12 pages)
- LC-110/H (50 pages)
- BETA LDT-500 (2 pages)
- M2000 (28 pages)
- MC1010 (46 pages)
- MC1210 (56 pages)
- TC100 (16 pages)
- T150 (8 pages)
- T140 Series (12 pages)
- PTC8010 (36 pages)
- PSC4010 (36 pages)
- MS-420 (1 page)
- MECP500 (2 pages)
- MECP2000 (2 pages)
- MECP10K (4 pages)
- MECP100 (2 pages)
- T130 (2 pages)
- T120 (2 pages)
- MC1200 (50 pages)
- MC1000 (28 pages)
- LC100 (20 pages)
- DMC1400 (106 pages)
- 320 (60 pages)
- 230 (106 pages)
- 130 (85 pages)
- QM100V (1 page)
- QM100V-40 (1 page)
- QM110T (2 pages)
- QM120ET (2 pages)
- QM130M (2 pages)