Martel Electronics MasterCAL User Manual
Page 91
Calibrator
Calibrating a Process Instrument
81
16.
Set
Trip Function
to
Trip Cont
by cycling through
the choices with the
e
key.
17.
Press the
Done
softkey.
18.
Press the
Manual Test
softkey.
19.
Close the pressure line vent and slowly bring the
pressure up to the trip point.
20.
When the switch sets, slowly bring the pressure
back down until the switch resets. You can
repeat this cycle as many times as you want.
21.
Press the
Done
softkey and view the results.
22.
Press the
Done
softkey and if desired, enter
Tag,
S/N,
and/or
ID
.
23.
Press the
Done
softkey.
24.
Exercise the switch by varying the applied
pressure. Adjust the switch until the set point is
correct.
25.
Press the
Done
softkey.
26.
Press the
As Left
softkey to run the test again
with the same parameters. Results from the As
Found and As Left tests are saved in the
calibrator memory for later viewing or uploading.
The procedure for limit switches that respond to
other parameters work similarly. When you do a 2 Pt.
Limit Switch Test, you simply follow the prompts on
the display for testing the first switch, changing test
leads, and testing the second limit switch.