AC Voltage Measurement

Frequency Range

% of Reading + Number of Counts

1 Year

2 Year

20 Hz to 40 Hz

2% + 10

2% + 10

40 Hz to 500 Hz

0.5% + 5

0.5% + 5

500 Hz to 1 kHz

2% + 10

2% + 10

1 kHz to 5 kHz

10% + 20

10% + 20

Ranges : 1.1000 V, 11.000 V, 110.00 V, 300.0 V rms

Resolution: 11.000 counts in all ranges except 300 V; 3,000 counts on 300 V range.

Input Impedance: 5 M

and <100 pF

Temperature Coefficient: 10% of specification/

°

C in the ranges -10 to 18

°

C and 28 to 50

°

C

Input Coupling: ac

Maximum Input Voltage: 300 V rms

Minimum Input Voltage: 0.5 V above 1 kHz

Specifications apply for 10% to 100% of voltage range.