Continuity testing, Frequency measurement – Martel Electronics MasterCAL User Manual
Page 120
MasterCAL
Users Manual
110
Continuity Testing
Tone Resistance
Continuous tone
May or may not get tone
No tone
<25
Ω
25 to 400
Ω
>400
Ω
Frequency Measurement
Ranges
Accuracy
1 Year
2 Year
1.00 Hz to 109.99 Hz
0.05 Hz
0.05 Hz
110.0 Hz to 1099.9 Hz
0.5 Hz
0.5 Hz
1.100 kHz to 10.999 kHz
0.005 kHz
0.005 kHz
11.00 kHz to 50.00 kHz
0.05 kHz
0.05 kHz
Minimum Amplitude for Frequency Measurement (square wave):
<1 kHz: 300 mV p-p
1 kHz to 30 kHz: 4 mV p-p
>30 kHz: 2.8 V p-p
Maximum input:
<1 kHz: 300 V rms
>1 kHz: 30 V rms
Input Impedance: 5 M
Ω
For frequency measurement less than 109.99 Hz, specifications apply for signals with a slew rate greater than 5
volt/millisecond.
See also other documents in the category Martel Electronics Equipment:
- 3001 (94 pages)
- BetaGauge 301 (40 pages)
- BetaGauge 311A-Ex (32 pages)
- BetaGauge 311A (48 pages)
- BetaGauge 330 (48 pages)
- BetaGauge 3II (197 pages)
- BetaGauge PI (20 pages)
- BetaGauge PIR (24 pages)
- BetaLOG-PI (28 pages)
- BetaLOG-TI (16 pages)
- BetaProbe TI/TI+ (9 pages)
- DPC-30 (12 pages)
- DPC300A (48 pages)
- DMC1410 (108 pages)
- IVC-222HPII (12 pages)
- LC-110/H (50 pages)
- BETA LDT-500 (2 pages)
- M2000 (28 pages)
- MC1010 (46 pages)
- MC1210 (56 pages)
- TC100 (16 pages)
- T150 (8 pages)
- T140 Series (12 pages)
- PTC8010 (36 pages)
- PSC4010 (36 pages)
- MS-420 (1 page)
- MECP500 (2 pages)
- MECP2000 (2 pages)
- MECP10K (4 pages)
- MECP100 (2 pages)
- T130 (2 pages)
- T120 (2 pages)
- MC1200 (50 pages)
- MC1000 (28 pages)
- LC100 (20 pages)
- DMC1400 (106 pages)
- 320 (60 pages)
- 230 (106 pages)
- 130 (85 pages)
- QM100V (1 page)
- QM100V-40 (1 page)
- QM110T (2 pages)
- QM120ET (2 pages)
- QM130M (2 pages)