FIas leftfl test run, As left” test run – Martel Electronics MasterCAL User Manual
Page 87
Calibrator
Calibrating a Process Instrument
77
4.
Adjust the transmitter output for 20 mA.
5.
If the span was adjusted in step 4, you must go
back and repeat steps 3 and 4 until no more
adjustment is required.
6.
Now check the transmitter at 50%. If it is within
specification, your adjustment is complete. If not,
adjust the linearity and begin this procedure
again at step 3.
“As Left” Test Run
Proceed as follows to generate and record
as left
data for the thermocouple temperature transmitter
you have just adjusted.
1.
Press the
As Left
softkey to record
as left data.
2.
Press the
Auto Test
softkey to begin an
automatic sequence through all the test points,
or you can step through the tests manually.
3.
When the tests are complete, observe the error
summary table, such as the following.
gj48s.eps
An asterisk (
*
) next to a measure or source value
indicates an unsettled value (
g
annunciator)
when the measurement was taken.
4.
If all the results are within specification, as they
are this time, press the
Done
softkey. An entry in
memory is made for
as left data.