If the process instrument needs more time to
settle than the calibrator’s normal settling time
(about 2 seconds) at each new stimulus level,
enter that time in seconds for

Delay

.

12.

Use the arrow keys to move the cursor down to
enter

0%

and

100%

values for SOURCE

temperature. Our example uses 100°C and
300°C.

13.

If the instrument calibration procedure requires
you to enter either the measurement value or
source value by hand at each step, press the

User Value

softkey, for “user entered values.”

Custom Units

lets you define your own units

such as “PH.” See “Creating Custom
Measurement Units,” earlier in this manual for an
example.

When you use custom units, the

w

symbol

appears next to the value on the display and in
results.

Press the

Done

softkey after you have

programmed your custom unit.

14.

The

Strategy

is the number of test points and

which test points are performed rising and falling
in percent of scale. Our example uses five points
(0%, 25%, 50%, 75%, and 100%), rising only.
Rising is indicated by the up arrow on the
display. Change to another test strategy by
pressing

e

on this line. A list of strategies

appears from which to choose. Select one, then
press

Done

.

15.

When you finish selecting the calibration
parameters, the display appears as follows:

gj45s.eps