beautypg.com

Dc voltage output – Martel Electronics MasterCAL User Manual

Page 121

background image

Calibrator

Specifications

111

DC Voltage Output

Range

Resolution

% of Output + % of Full Scale

1 Year

2 Year

110 mV

1

µ

V

0.01% + 0.005%

0.015% + 0.005%

1.1 V

10

µ

V

0.01% + 0.005%

0.015% + 0.005%

15 V

100

µ

V

0.01% + 0.005%

0.015% + 0.005%

Temperature Coefficient: (0.001% of output + 0.001% of f.s.)/

°

C in the ranges -10 to 18

°

C and 28 to 50

°

C

Maximum Output Current: 10 mA

Loading: (0.001% f.s. + 1

µ

V)/ mA

Common Mode Error: 0.008% f.s/(Common Mode Volt)

Maximum Input Voltage: 30 V dc