9.7

Fault finding

Most of the common faults and symptoms experienced with the flowmeter can be resolved
with the help of the following table.
To simplify the use of the table, the faults and symptoms are grouped together:

GROUPS

D

Display, inputs and outputs

I

Current output

P

Pulse output

A

Alarm output (Status)

E

Control input (Binary)

OP

Measurement mode and commissioning

ST

Commissioning and Start-up of the flowmeter

Group

Fault / Symptom

Cause

Remedy

Group D

D1

No display or outputs

Power not switched on

Switch on power

Power supply fuse F9
blown

Replace fuse as per
Section 9.3.1

Fuses F10, and/or
F12 blown

Replace converter as per
Sect. 9.2 or call Krohne
service.

D2

Fluctuating display and
outputs

Time constant too small

Increase time constant as
per Section 6.3.

D3

Mass flow display
incorrect

The wrong value for
parameters CF5 (GK)
programmed. (This is
stamped on the data plate)

Check and correct
according to
Sections 6.12 and 6.14.4

Zero calibration

Re-do zero, check manual
offset

Primary sensor faulty

Check as per Section 9.8

D4

Density display and
outputs not correct

Density Parameters

Check and correct
according to 6.12

Excitation frequency of
primary sensor not correct
when filled with water (see
section 2.7.3)

Check for air in meter.
Call Krohne.

Primary sensor faulty

Check as per Section 9.8

Group I

I1

Connected instrument
displays 0 or negative
values

Connection polarity
wrong

Correct as per Sect. 3.3

Connected instrument
faulty or current output
faulty

Check output with a mA
meter.
I Test OK
Check cabling and
connected instrument and
replace if necessary.
I Test faulty
Current output faulty.
Replace converter or call
Krohne service.

Please check the following table of hints and advice
before calling your KROHNE Service Department.