beautypg.com

KROHNE CORIMASS E Series EN User Manual

Page 52

background image

52

To set the Time Constant :

Begin from measuring mode.

This filtering only applies to mass and volume flow readings and any outputs that use them.
The mass totaliser is independent of the time constant.

The standard range for the time constant is 0.5 to 20 seconds. For faster response times, a
range of 0.2 to 20 seconds is available as a factory set option.

Fig. 33: Time constant characteristic

6.4

Programming the display for measurement values (Fct. 1.2. and 3.2.)

The following measurement functions can be displayed :

For systems with concentration software their specific extra displays will be added to the
above list from Fct. 1.2.9 onwards.

Key

Display
line 1

line 2

→→→

Fct. 1.1.(1)

ZERO SET

↑↑

Fct. 1.1.(3)

TIME. CONST.

(0)4.0

TIME C. S.

Edit the time constant in
seconds in range 0.5 to 20.

Fct. 1.1.(3)

TIME.CONST

4x

Fct. 1.2.1

CYCL.DISP

Fct. 1.2.2

STATUS MSG.

Fct. 1.2.3

MASS FLOW

Fct. 1.2.4

MASS TOTAL

Fct. 1.2.5

DENSITY

Fct. 1.2.6

TEMPERAT.

Fct. 1.2.7

VOLUME.FLOW

Fct. 1.2.8

VOL.TOTAL

This manual is related to the following products: