beautypg.com

KROHNE CORIMASS E Series EN User Manual

Page 75

background image

75

The default value of the custody transfer protection code when leaving the factory is

↵→↑↵↑→↵→↑

This password may be changed using Menu 3.8.7.. However, this can only be done if custody
transfer protection is disabled first, as described above.

Begin from measuring mode.

If the two entered passwords are different the message "CODE WRONG" is displayed.

The message has to be acknowledged with the

key and input must be repeated with Fct.

3.8.7.. Then the protection state 'active' or 'inactive' can be chosen with Fct.3.8.6..

NOTE:
If the input of the custody transfer password is wrong, a 9-character code is displayed. With
this character code the manufacturer can decode the password in cases where the password
is lost.

It is also possible to protect just the mass totaliser. Menu 3.8.5. ENABL.RESET determines
whether the operator can clear the mass total from the ACKNOWLEDGE/RESET MENU.

Begin from measuring mode.

Key

Display
line 1

line 2

Fct. (1).0

OPERATOR

2x

Fct. (3).0

INSTALL.

Fct. 3.(1).0

BASE PARAM.

7x

Fct. 3.(8).0

USER DATA

Fct. 3.8.(1)

LANGUAGE

6x

Fct. 3.8.(7)

CODE 3

CodeE 3

- - - - - - - -

Enter new custody transfer
password two times.
Fct. 3.8.(7)

CODE 3

4x

Key

Display
line 1

line 2

Fct. (1).0

OPERATOR

2x

Fct. (3).0

INSTALL

7x

Fct. 3.(8).0

USER DATA

4x

Fct. 3.8.(5)

ENABL.RESET

(YES)

(NO)

Fct. 3.8.(5)

ENABL.RESET

4x

+110.25

kg

Mass Total Display

CodE 2

- -

↑→

RESET MASS

BLOCKED

Totaliser Reset is disabled

↵↵

This manual is related to the following products: