Part a system installation and start-up, 1 installation in the pipeline – KROHNE BATCHCONTROL 5014 C User Manual
Page 6

Installation and operating instructions
6 BATCHCONTROL
 
 
Part A System installation and start-up
 
 
1 Installation in the pipeline
1.1 Important
information
 
The following recommendations should be observed to ensure proper functioning of the 
flowmeter – PLEASE NOTE. 
 
•
Measuring tube must be filled 
completely at all times. 
•
Direction of flow:  the blue arrow on the 
housing of the primary head must point in 
the direction of flow. If for structural 
reasons the flowmeter can only be installed 
invers to the direction of flow, the direction 
of flow measurement can be reprogram-
med. 
•
Stud bolts and nuts: to fit, make sure 
there is sufficient room next to the pipe 
flanges. 
•
Vibration: support the pipeline on both 
sides of the flowmeter. Vibration level to 
IEC 068-2-34: below 2,2g in the 
20 - 2000 Hz frequency range. 
•
Radiant heat: avoid e.g. from hot product 
tanks, insulate if necessary. 
•
Avoid strong electromagnetic fields 
in vicinity of flowmeter. 
•
Straight Inlet run
≥ 5 x DN and straight
outlet run
≥ 2 x DN, measured from the
electrode axis (DN = meter size).
•
Vortex or corkscrew flow: increase 
length of inlet and outlet runs or install flow 
straighteners. 
•
Mixing different process liquids:  install 
flowmeter upstream of mixing point or at an 
adequate distance downstream, minimum 
of 30 x DN (DN = meter size), otherwise 
display may be unsteady. 
•
Plastic pipes and internally coated metal 
pipes: grounding rings required, see 
“Grounding“, Section 1.3.3. 
•
Heat-insulated pipelines: do not insulate 
flowmeter. 
•
Zero setting: not required. For checking 
purpose, it should be possible to set “zero“ 
flow velocity in the completely filled 
measuring tube. Shutoff valves should 
therefore be provided either downstream or 
upstream and downstream of the 
flowmeter. 
•
Ambient temperature 
-25°C to +60°C 
•
Process temperature 
max. 140 °C 
•
Transport and storage temperature 
-25°C to +60°C 
 
Limits imposed by the material used for the 
measuring tube for process temperature, 
thermal shock limit, pressure and vacuum, see 
Section D page 22. 
Please note!
The ceramic measuring tube must not contact 
metal parts (flange, pipeline). This can destroy 
the flowmeter! 
 
 
 
 
 
 
 
 
 
 
 
1