HARTÆ Communications Protocol

IFC010 Command Specification Rev: 1.0

Release Date: 29 May 1997

© 1997 KROHNE Mefltechnik D-47058 Duisburg

Page

26

0

No Command-Specific Errors

1-4

Undefined

5

Too Few Data Bytes Received

6-7

Undefined

8

Update Failure

9-14

Undefined

15

Wrong Password

16-127

Undefined

8.20

Command #151

Control Zero Calibration

Unlike Command #43, Set Primary Variable Zero, where an instant measurement is used to correct zero point, this
command performs calibration exactly as if this process was activated via device keypad (Menu 3.3): 255 subsequent
measurements are averaged before zero correction is done. The process lasts 3.5-5.5 sec (best and worse cases).

REQUEST DATA BYTES

#0
CAL
CONTROL
VAR

RESPONSE DATA BYTES

#0

#1

CAL

CAL

CONTROL

STATUS

VAR

Data Byte #0

Calibration control variable:
0 - Exit calibration
1 - Start calibration
2 - Store calibration value
3 - Get calibration status

Data

Byte

#2

Calibration status corresponds to the number of iterations carried out. After 255
measurements are ìcollectedî, calibration as a process is finished, and calculated zero is
preserved in RAM until Command #151 will be received with StartCalibration or
StoreCalValue value of control variable.

NOTE: After calibration is activated, any subsequent StartCalibration request is ignored; for such requests zero is
responded as CalibrationStatus. Until StartCalibration request is not concluded by either ExitCalibration or
StoreCalValue, it will not be possible to start another calibration.

COMMAND-SPECIFIC RESPONSE CODES

0

No Command-Specific Errors

1

Undefined

2

Invalid Selection (Of Control Variable)

3-4

Undefined

5

Too Few Data Bytes Received

6

Undefined

7

In Write Protect Mode

8

Update Failure

9

No Zero Value (Iterationsí Number < 255)

10-127

Undefined

8.21

Command #152

Quit Errors