8.20
Command #151
Control Zero Calibration
Unlike Command #43, Set Primary Variable Zero, where an instant measurement is used to correct zero point, this
command performs calibration exactly as if this process was activated via device keypad (Menu 3.3): 255 subsequent
measurements are averaged before zero correction is done. The process lasts 3.5-5.5 sec (best and worse cases).
REQUEST DATA BYTES
#0
CAL
CONTROL
VAR
RESPONSE DATA BYTES
#0
#1
CAL
CAL
CONTROL
STATUS
VAR
Data Byte #0
Calibration control variable:
0 - Exit calibration
1 - Start calibration
2 - Store calibration value
3 - Get calibration status
Data
Byte
#2
Calibration status corresponds to the number of iterations carried out. After 255
measurements are ìcollectedî, calibration as a process is finished, and calculated zero is
preserved in RAM until Command #151 will be received with StartCalibration or
StoreCalValue value of control variable.
NOTE: After calibration is activated, any subsequent StartCalibration request is ignored; for such requests zero is
responded as CalibrationStatus. Until StartCalibration request is not concluded by either ExitCalibration or
StoreCalValue, it will not be possible to start another calibration.
COMMAND-SPECIFIC RESPONSE CODES
0
No Command-Specific Errors
1
Undefined
2
Invalid Selection (Of Control Variable)
3-4
Undefined
5
Too Few Data Bytes Received
6
Undefined
7
In Write Protect Mode
8
Update Failure
9
No Zero Value (Iterationsí Number < 255)
10-127
Undefined
8.21
Command #152
Quit Errors