HARTÆ Communications Protocol

IFC010 Command Specification Rev: 1.0

Release Date: 29 May 1997

© 1997 KROHNE Mefltechnik D-47058 Duisburg

Page

25

0

No Command-Specific Errors

1

Undefined

2

Invalid Selection (of Language)

3-4

Undefined

5

Too Few Data Bytes Received

6

Undefined

7

In Write Protect Mode

8

Update Failure

9-127

Undefined

8.19

Command #150

Enable/Disable Password Protection

Installs/uninstalls write protection. Two transaction are implemented: Transaction #1 disables write protection, in this
case a correct password must be ìshownî to transmitter. If protection is disabled and device is open for write operations,
any of two transactions can be used to enable factory-set password. Then the contents of Password field for Transaction
#1 is ignored.

TRANSACTION #0

REQUEST DATA BYTES

#0
WRITE
PROT
CODE

RESPONSE DATA BYTES

#0
WRITE
PROT
CODE

TRANSACTION #1

REQUEST DATA BYTES

#0

#1

#2

#3

#4

WRITE

PROT

PROT

PROT

CODE

CODE

CODE

MSB

LSB

RESPONSE DATA BYTES

#0
WRITE
PROT
CODE

Data Byte #0

Write Protect Code, available via Command #15, Read Primary Variable Output Information:

0 - Not Write Protected
1 - Write Protected

Data Byte #1-#4

HART- encrypted password of the instrument, unsigned long. The encoding scheme is as
follows: the ìRightî-key of device keypad is coded by digit м0о, ìEnterî-key - by м1о and
ìUpî-key - by м2о. For example, some arbitrary key combination ìRight-Right-Up-Up-Up-
Right-Enter-Up-Upî will result in 2220122 (leading zeroes are defaulted for password
operations anywhere).

NOTE: Any nonzero value of Write Protect Code will be interpreted by transmitter as a request to enable protection.

COMMAND-SPECIFIC RESPONSE CODES