8 command #137 reset totalizers, 9 command #138 read analog output function – KROHNE IFC 010 Converter HART User Manual
Page 17
HARTÆ Communications Protocol
IFC010 Command Specification Rev: 1.0
Release Date: 29 May 1997
© 1997 KROHNE Mefltechnik D-47058 Duisburg
Page
17
0
No Command-Specific Errors
1-4
Undefined
5
Too Few Data Bytes Received
6
Undefined
7
In Write Protect Mode
8
Update Failure
9-127
Undefined
8.8
Command #137
Reset Totalizers
Clears (sets to zero) both totalizers.
REQUEST DATA BYTES
NONE
RESPONSE DATA BYTES
NONE
COMMAND-SPECIFIC RESPONSE CODES
0
No Command-Specific Errors
1-6
Undefined
7
In Write Protect Mode
8
Update Failure
9-15
Undefined
16
Access Restricted
17-127
Undefined
8.9
Command #138
Read Analog Output Function
Reads the function of current/pulse output.
REQUEST DATA BYTES
#0
ANALOG
OUTPUT
NUMBER
CODE
RESPONSE DATA BYTES
#0
#1
ANALOG
ANALOG
OUTPUT
OUTPUT
NUMBER
FUNCTION
CODE
CODE
Data Byte #0
Analog Output Number Code, equals м1о for current output and м2о - for p ulse output
Data Byte #1
Analog Output Function Code:
0 - Off,
1 - 1 Dir, output traces flowrate measurements only in one direction,
3 - 2 Dir, output traces flowrate measurements in both directions.
COMMAND-SPECIFIC RESPONSE CODES
0
No Command-Specific Errors
1
Undefined
2
Invalid Selection