HARTÆ Communications Protocol
IFC010 Command Specification Rev: 1.0
Release Date: 29 May 1997
© 1997 KROHNE Mefltechnik D-47058 Duisburg
Page
22
Data Byte #0
0 - ì+ Dirî (corresponds to the arrow direction on primary head)
1 - ì- Dirî
Data Byte #1-#4
Primary head constant GKL, IEEE 754.
COMMAND-SPECIFIC RESPONSE CODES
0
No Command-Specific Errors
1-127
Undefined
8.16
Command #147
Write Primary Constant And Flow Direction
Writes the primary head constant GKL and flow direction.
REQUEST DATA BYTES
#0
#1
#2
#3
#4
FLOW
PRIMARY
PRIMARY
DIRECTION
HEAD
HEAD
VALUE
VALUE
MSB
LSB
RESPONSE DATA BYTES
#0
#1
#2
#3
#4
FLOW
PRIMARY
PRIMARY
DIRECTION
HEAD
HEAD
VALUE
VALUE
MSB
LSB
Data Byte #0
0 - ì+ Dirî (corresponds to the arrow direction on primary head)
1 - ì- Dirî
Data Byte #1-#4
Primary head constant GKL, IEEE 754. Valid settings: 1.0000 ... 9.9999 (refer to primary
head nameplate).
COMMAND-SPECIFIC RESPONSE CODES
0
No Command-Specific Errors
1
Undefined
2
Invalid Selection (of FlowDirection)
3
Passed Parameter too Large (PrimaryHeadValue)
4
Passed Parameter too Small (PrimaryHeadValue)
5
Too Few Data Bytes Received
6
Undefined
7
In Write Protect Mode
8
Update Failure
9-127
Undefined
8.17
Command #148
Read Language And User-Defined Unit
Reads data defining user-created unit, and language that is used for text presentation on the local screen of the
instrument.
REQUEST DATA BYTES
NONE
RESPONSE DATA BYTES
#0
#1
...
#5
DISPLAY
VOLUME
VOLUME