Fct. 3.3.9 Cutoff “off” value for SMU-I
(appears only if “YES” set in fct. 3.3.7).
Range: 02 to 20 PERCENT of Q

100%

.

This value must be greater than that of fct. 3.3.8,
otherwise a “parameter check” error (fct. 4.5.0) will
occur, see sect. 4.2 + 4.3. When the flow returns to the
cutoff “off” value, the current output returns to normal.

Fct. 3.4.6 Low-flow cutoff (SMU) for F desired?
Set NO or YES.

Fct. 3.4.7 Cutoff “on” value for SMU-F
(appears only if “YES” is set in fct. 3.4.6).
Range: 01 to 20 PERCENT of Q

100%

.

The low-flow cutoff drives the frequency output to 0 Hz
when the flow decreases to the cutoff “on” value.

Fct. 3.4.8 Cutoff “off” value for SMU-F
(appears only if “YES” is set in fct. 3.4.6).
Range: 02 to 20 PERCENT of Q

100%

.

T h is valu e m u st be greater th an th at of fct. 3 .4 .7 , oth erw ise a
“ param eter ch eck” error (fct. 4 .6 .0 ) w ill occu r, see sect. 4 .2 +
4 .3 . W h en th e flow retu rn s to th e cu toff “ off” valu e, th e
freq u en cy ou tpu t retu rn s to n orm al.

5.11

F/R operation for I or F

Electrical connection, characteristic and setting of
outputs, refer to sect. 2.3, 5.7 + 5.8.
Fct. 3.1.7 Define direction of forward (normal) flow
(+ or -).
In the case of F/R operation, set the direction of the
forward flow with “+” or “-”, in accordance with the
arrows marked “+” and “-” on the primary head. Note
that if the flowmeter does not have a “+” and “-” on the
primary head the “+” direction is the direction in which
the arrow is pointing. When the low-flow cutoff is active,
it will also operate with the F/R operation.

Fct. 3.1.1 Full-scale setting for flowrate Q

100%

Set the full-scale range. Unit and range, see sect. 5.1 +
5.3.

Fct. 3.1.2 Separate range desired for reverse flow?
Set “YES” only if the range required for reverse flow is
different from the normal (forward) flow. If not, set
NO”.

Fct. 3.1.3 Full-scale range for reverse flow
(appears only if “YES” is set in fct. 3.1.2).
This is used for setting the full-scale range for reversed
flow. Unit and range, see sect. 5.1 + 5.3. This value must
not be greater than that of fct. 3.1.1, otherwise
“parameter check” error (fct. 4.2.0) will occur, see sect.
4.2 + 4.3.

5.12

Language of display texts

A choice of languages for other display texts is offered in
fct. 3.6.1. The software is equipped with the languages:

GB/USA

English and

D

German or

F

French.

Other languages can be delivered on request.

5.13

Coding desired for entry into setting level?

Set NO or YES in fct. 3.6.2.

If NO is set, only press the

key to get into the

setting mode.

If YES is set, press the

key and then a 9-keystroke

combination to enter the setting mode.

Factory-set Entry Code 1:

To change the Entry Code 1:

Select fct. 3.6.2 ENTRY.CODE 1: set YES.
Select fct. 3.6.3 CODE 1 (appears only if YES is set
in fct. 3.6.2). Press the

key, displayed:

Code 1 _ _ _ _ _ _ _
Press any 9-keystroke combination. Every keystroke
acknowledged by “*”. Then set the same keystroke
combination again. WRONG CODE (= incorrect
entry) appears if 1

st

and 2

nd

entries are not equal.

Press the

and the

keys and repeat settings.

5.14

Behavior of outputs during setting

Behaviour of outputs during setting Set in fct. 3.6.5
whether or not the outputs shall hold the last values
(before entry into the setting mode).

If “YES” is set: output values prior to entry into the
setting mode are retained during setting. After
leaving the setting mode, the outputs go to the values
corresponding to the actual operating conditions.

If “NO” is set: the outputs go to the set minimum
values:

I

to 0/4 mA (see fct. 3.3.2)
or to the value of I

0%

(see fct. 3.3.3).

F

or to 0 V, consequently no pulses.

5.15

User-defined unit

An arbitrary volumetric flow unit or, if density of the
fluid product is consistent and known, a unit of mass
(weight) can be set in fct. 3.6.6 to 3.6.8. The unit “ h
Liter
/hr” (hecto-liters per hour) is factory-set unless
another special unit is specified. US-version: “US
Mgal/DAY
” (US million gallons per day).

Fct. 3.5.6 Text for user-defined unit

Volumetric (or mass) unit per unit of time.

Text for volume (mass): 6 characters (places).

Text for time: 3 characters (places).

The fraction bar “/” in the 7

th

place has a fixed

position.

Alpha characters A-Z and a-z, numbers 0-9, symbols
+, - or blank character (= underscore) are selectable
for every place.

Pressing the

key will sequence the alpha characters

and numbers in the order given above.

The

key shifts the cursor 1 place to the right.

Text examples are given in the following tables in
brackets (...../..).