Changing the numbers.

4.2.4

Example for setting the signal converter

Display

Keys

1

st

line

2

nd

line

Explanation

6.0000

E

1

PuLSe/min

3.6000

E

3

PuLSe/hr

3.6000

E

3

PuLSe/hr

8 x

1.6000

E

3

PuLSe/hr

1.6000

E

3

PuLSe/hr

4 x

1.0000

E

3

PuLSe/hr

Change pulse rate
(Continuation)

Fct. 3.4.3

PULSRATE

Fct. 3.4

FREQ.OUTP.F

Fct. 3.0

INSTALL.

STORE YES

Return to measuring
mode

PARAM.CHEC
K

If fct. 4.0 PARAM.ERR.
should appear:
plausibility error, see
sect. 4.3

- - - - - - -

- - - - - - - / - - -

Measuring mode

Display

Keys

1

st

line

2

nd

line

Explanation

- - - - -

- - - - - - -/ - -

Measuring mode

Fct. 1.0

OPERATION

When ENTRY CODE 1
“YES” is set in fct. 3.6.2
the 9 key-stroke “CodE 1”
must be typed in.
Factory setting

Main menu operation

2 x

Fct. 3.0

INSTALL.

Main menu setting

Fct. 3.1

BASIS
PARAM

Submenu Frequency output

3 x

Fct. 3.4

FREQ.OUTP.
F

Fct. 3.4.1

FUNCTION F

2 x

Fct. 3.4.3

PULSRATE

Change pulse rate

1.0000 E

PuLSe/Sec

Changing the texts (units).
When the units are changed, the totalizer value is automatically converted.

1 3 . 5 7 1

m

3

/ h r

9 7 . 3 5

m

3

/ h r

9 7 . 3 5

m

3

/ h r

Moving the cursor (blinking part) to the

i ht

9 7 3 5 0

L i t e r / h r

1 3 . 5 7 1

m

3

/ h r

Changing from text (unit) to totalizer setting.

Changing from totalizer setting back to text.

1 0 . 3

S e c

9 7 . 4 5

m

3

/ h r

9 7 . 3 5

m

3

/ h r

7 . 1 4 3 7

U S . G a l / S e c

1 3 . 5 7 1

m

3

/ h r

1 3 . 5 7 1

m

3

/ h r

Fc t . 1 . 3 . 1

T I M E C O N S T . I

Return to display of functions.

The following example shows how to change the pulse rate of the frequency output (function: fct. 3.4.3 PULSRATE, see
sect. 4.3). The cursor (flashing part of display) is shown here in “bold type.
- “old” setting:

1 pulse per second

(1.000 E 0 PuLSe/Sec)

- change to:

1000 pulses per hour

(1.000 E 3 PuLSe/hr)