Frequency output for flowrate Q = 100% (full-scale range) for F/R operation and setting in PULSE/UNIT (fct. 3.4.2 +
3.4.3).

T h e freq u en cy at th e ou tpu t alw ays refers to th e settin g of th e forw ard ran ge F

100%

(in case of forw ard flow ) or to th e reverse ran ge

R

100%

(in case of reverse flow ).

5 .9

S tatus outp ut S

5 .9 .1

A p p lication S (fct. 3 .5 .1 )

A p p lication S

S etting via fct. 3 .5 .1

F atal error

F A T A L E R R .

U ltrason ic error

U S E R R O R

F orw ard /R everse in d ication

F /R IN D .

T rip poin t

T R IP P O IN T

(N ote: w h en th e in d ication ou tpu t con tact is closed , th e open collector ou tpu t is active.)
F or con n ection d iagram s of th e statu s ou tpu t, see sect. 2 .3 .5 .
F atal error

- D ata error in E E P R O M 1 (param eters).
- C alibration d ata lost. (E E P R O M 2 ).
- C u rren t calibration valu es.
- T otalizer cou n ts lost or totalizer overflow .
- R A M error.
- R O M error.
- F overran ge (1 2 5 % ).
- I overran ge (1 2 5 % ).
- E m pty pipe.

C on tact w ill be open w ith F atal error an d closed w h en n o error occu rs.
U S error

O n e or both u ltrason ic path s fail. C on tact w ill be open in case of path failu re an d closed w h en n o path
failu re.

F /R ind .

F low d irection in d ication poin t 1 is u sed as h ysteresis in percen tage of th e forw ard fu ll-scale. C on tact w ill
be open d u rin g forw ard flow an d closed d u rin g reverse flow .

T rip p oint

P ossibility for open in g or closin g statu s ou tpu t con tact, w h en flow ex ceed s a pre-d efin ed poin t.

If (p oint 1 > p oint 2 )

T h e con tact w ill close if flow rate rises above poin t 1 an d it w ill open if flow rate d rops below poin t 2 .

If (p oint 2 > p oint 1 )

T h e con tact w ill open if flow rate rises above poin t 2 an d it w ill close if flow rate d rops below poin t 1.

5.10

Low-flow cutoff (SMU) for I + F

T o avoid erron eou s m easu rem en ts at low flow rates, th e
S M U sw itch es th e cu rren t an d freq u en cy ou tpu ts (I+ F ) off.
I goes to 0 /4 m A (fct. 3 .3 .2 ) or to I

0%

(fct. 3 .3 .3 ) an d F to 0

H z.

If “ N O ” is set u n d er fu n ction s 3 .3 .7 + 3 .4 .6 , fixed cu toff
“ on ” an d cu toff “ off” valu es of 0 .1 an d 0 .2 5 % , resp., of
Q

100%

(fu ll-scale ran ge, see fct. 3 .1 .1 ) act on th e ou tpu ts I +

F .

If “ YE S ” is set u n d er fu n ction s 3 .3 .7 + 3 .4 .6 , th e cu toff
“ on ” an d cu toff “ off” valu es for I + F are separately
ad ju stable in th e ran ges specified below .

F ct. 3 .3 .7 L ow -flow cutoff (S M U ) for I d esired ?

S et N O or YE S .

F ct. 3 .3 .8 C utoff “ on” value for S M U -I
(appears on ly if “ YE S ” set in fct. 3 .3 .7 ).

R an ge: 0 1 to 1 9 P E R C E N T of Q

100%

.

T h e low -flow cu toff d rives th e cu rren t ou tpu t to 0 /4 m A (fct.
3 .3 .2 ) or to I

0%

(fct. 3 .3 .3 ) w h en th e flow d ecreases to th e

cu toff “ on ” valu e.

F

F

max

(125%)

F

100%

V

max

Max

0

Sound velocity measurement

V

min

V

sound

(m/s)

F5

100 Max %

%

0

100

Max

Contact open

Switched off

F

F4

Q

forwar

Q

reverse