4.3

Error messages

4.3.1

Description of error messages as shown in display

The following list gives all errors that can occur during process flow measurement.

List of
errors

*

When errors are displayed during the measuring mode and in the ERRORLIST in the Reset/Quit menu, “a
numeral” and “Err” will appear in the 1

st

(top) line. The numeral gives the number of momentarily occurring

errors that are displayed alternately with the actual measured value.

**

No output in measuring mode. With these errors, the signal converter is automatically in the setting mode.

***

These errors are additionally identified by the flashing compass field.

Invoke and then terminate setting mode

.

Keystrokes:

and

or

/ [9-keystroke entry code 1] and

(dependent on setting in fct. 3.6.2); then

acknowledge error(s) in Reset/Quit menu.

Invoke and then terminate setting mode.
Keystrokes:

and

or

/ [9-keystroke entry code 1] and

(dependent on setting in fct. 3.6.2); then

acknowledge error(s) in Reset/Quit menu.
Consult factory if these errors occur several times in succession.

Check electrical connection of US path 1+2 and make sure that the measuring tube is completely filled with fluid.
If in order, please consult factory.

2 Err

A

D

B

D

C

Error messages

Error output in measuring mode via display

(see fct. 3.2.6) dependent on setting.

Display

2

nd

(middle) line

Description of error

Rectify instrument fault

and / or

clear error message

NO MESS

US

ERROR

TOTAL

ERROR

ALL

ERROR

LINE INT.

Power failure since last setting
Note: no counting during power failure


Reset totalizer(s) if necessary

-

-

yes

yes

TOTALIZER

Counts lost or totalizer overflow
Note: totalizer was reset!

-

-

yes

yes

EEPROM 2

Data error in EEPROM 2 (totalizer)
Note: totalizer deviation possible


Reset totalizer(s) if necessary

-

-

yes

yes

RAM

Check-sum error in RAM

-

-

-

yes

ROM

Check-sum error in ROM

-

-

-

yes

US PATH 1 ***

US path 1 faulted

-

yes

-

yes

EMPTY PIPE ***

Measuring tube empty

-

yes

-

yes

US PATH 2 ***

US path 2 faulted

-

yes

-

yes

FREQ.OUTP. F

Frequency output overranged


If necessary check data, fct.3.4.0

-

-

-

yes

CUR. OUTP. I

Current output overranged


If necessary check data, fct.3.3.0

-

-

-

yes

EEPROM 1

Data error in EEPROM 1
(parameters)

Check instrument parameters

**

**

**

**

CAL. DATA

Calibration data lost

Recalibrate signal converter, please
consult factory

**

**

**

**

EE1 EE2

Current calibration values EEPROM
1 + 2 are different

Terminate setting mode
(press 1 x

-key), values corrected

automatically.

**

**

**

**

Error representation

A

Compass field (flashes when errors marked “

∗∗∗

” above occur).

B

Number of errors that have occurred.

C

Plain text for error message(s).

D

With bar:
“new” errors, not yet acknowledged.
Without bar:
“old”, acknowledged errors.

see section 4.4.3