21
4.3
Error messages
4.3.1
Description of error messages as shown in display
The following list gives all errors that can occur during process flow measurement.
List of
errors
*
When errors are displayed during the measuring mode and in the ERRORLIST in the Reset/Quit menu, “a
numeral” and “Err” will appear in the 1
st
(top) line. The numeral gives the number of momentarily occurring
errors that are displayed alternately with the actual measured value.
**
No output in measuring mode. With these errors, the signal converter is automatically in the setting mode.
***
These errors are additionally identified by the flashing compass field.
❑
Invoke and then terminate setting mode
.
Keystrokes:
→
and
↵
or
→
/ [9-keystroke entry code 1] and
↵
(dependent on setting in fct. 3.6.2); then
acknowledge error(s) in Reset/Quit menu.
❍
Invoke and then terminate setting mode.
Keystrokes:
→
and
↵
or
→
/ [9-keystroke entry code 1] and
↵
(dependent on setting in fct. 3.6.2); then
acknowledge error(s) in Reset/Quit menu.
Consult factory if these errors occur several times in succession.
✧
Check electrical connection of US path 1+2 and make sure that the measuring tube is completely filled with fluid.
If in order, please consult factory.
2 Err
A
D
B
D
C
Error messages
Error output in measuring mode via display
(see fct. 3.2.6) dependent on setting.
Display
2
nd
(middle) line
Description of error
Rectify instrument fault
and / or
clear error message
NO MESS
US
ERROR
TOTAL
ERROR
ALL
ERROR
LINE INT.
Power failure since last setting
Note: no counting during power failure
❑
Reset totalizer(s) if necessary
-
-
yes
yes
TOTALIZER
Counts lost or totalizer overflow
Note: totalizer was reset!
❑
-
-
yes
yes
EEPROM 2
Data error in EEPROM 2 (totalizer)
Note: totalizer deviation possible
❑
Reset totalizer(s) if necessary
-
-
yes
yes
RAM
Check-sum error in RAM
❍
-
-
-
yes
ROM
Check-sum error in ROM
❍
-
-
-
yes
US PATH 1 ***
US path 1 faulted
✧
-
yes
-
yes
EMPTY PIPE ***
Measuring tube empty
✧
-
yes
-
yes
US PATH 2 ***
US path 2 faulted
✧
-
yes
-
yes
FREQ.OUTP. F
Frequency output overranged
❑
If necessary check data, fct.3.4.0
-
-
-
yes
CUR. OUTP. I
Current output overranged
❑
If necessary check data, fct.3.3.0
-
-
-
yes
EEPROM 1
Data error in EEPROM 1
(parameters)
Check instrument parameters
❍
**
**
**
**
CAL. DATA
Calibration data lost
Recalibrate signal converter, please
consult factory
**
**
**
**
EE1 EE2
Current calibration values EEPROM
1 + 2 are different
Terminate setting mode
(press 1 x
↵
-key), values corrected
automatically.
**
**
**
**
Error representation
A
Compass field (flashes when errors marked “
∗∗∗
” above occur).
B
Number of errors that have occurred.
C
Plain text for error message(s).
D
With bar:
“new” errors, not yet acknowledged.
Without bar:
“old”, acknowledged errors.
see section 4.4.3