KROHNE UFM 500 EN User Manual
Page 16

16
4.2.5
RESET / QUIT Menu, reset totalizers and delete error messages
Resetting totalizers
Displaying and deleting error messages
Keys
Display
Description
- - - - - - -
- - - - - - / - - -
Measuring mode.
↵
CodE 2
- -
Type in entry-code 2 for RESET/QUIT-menu:
↑
→
.
↑
→
TOTAL.RESET
Totalizer menu only appears when ENABL.RESET “YES” is set in
fct. 3.6.10, if not display will show “ERROR LIST”, see next section.
→
TOTAL. +
(
↑
)
(TOTAL. -)
If necessary, choose “-Totalizer” with
↑
key.
→
RESET NO
Do not reset totalizer, press
↵
3 times to return to measuring mode.
↑
RESET YES
Reset totalizer.
↵
TOTAL. + (-)
“+” (or “-“) totalizer has been reset.
If necessary choose the other totalizer with
↑
key and reset as well.
↵
TOTAL.RESET
↵
- - - - - - -
- - - - - - / - - -
Return to measuring mode.
Keys
Display
Description
- - - - - - -
- - - - - - / - - -
Measuring mode.
↵
CodE 2
- -
Type in entry-code 2 for RESET/QUIT-menu:
↑
→
.
↑
→
TOTAL.RESET
Totalizer menu only appears when ENABL.RESET “YES” is set in
fct. 3.6.10, if not display will show “ERROR.LIST”, see next section.
↑
ERROR.LIST
Menu for status messages.
→
- - - - - - - - - - -
1. Error message is displayed.
↑
- - - - - - - - - - -
2. Error message is displayed.
↑
,
↑
, ...
- - - - - - - - - - -
Further error messages are displayed, if present at the end of the list.
ERROR.LIST
ERRORLIST is displayed.
→
QUIT.NO
Do not delete error messages,
press
↵
key three times = return to measuring mode.
↑
QUIT.YES
Delete error messages.
↵
ERROR QUIT
Error messages have been deleted.
↵
ERROR.LIST
↵
- - - - - - / - - -
Return to measuring mode.