4.2.5

RESET / QUIT Menu, reset totalizers and delete error messages

Resetting totalizers

Displaying and deleting error messages

Keys

Display

Description

- - - - - - -

- - - - - - / - - -

Measuring mode.

CodE 2

- -

Type in entry-code 2 for RESET/QUIT-menu:

.

TOTAL.RESET

Totalizer menu only appears when ENABL.RESET “YES” is set in
fct. 3.6.10, if not display will show “ERROR LIST”, see next section.

TOTAL. +

(

)

(TOTAL. -)

If necessary, choose “-Totalizer” with

key.

RESET NO

Do not reset totalizer, press

3 times to return to measuring mode.

RESET YES

Reset totalizer.

TOTAL. + (-)

“+” (or “-“) totalizer has been reset.
If necessary choose the other totalizer with

key and reset as well.

TOTAL.RESET

- - - - - - -

- - - - - - / - - -

Return to measuring mode.

Keys

Display

Description

- - - - - - -

- - - - - - / - - -

Measuring mode.

CodE 2

- -

Type in entry-code 2 for RESET/QUIT-menu:

.

TOTAL.RESET

Totalizer menu only appears when ENABL.RESET “YES” is set in
fct. 3.6.10, if not display will show “ERROR.LIST”, see next section.

ERROR.LIST

Menu for status messages.

- - - - - - - - - - -

1. Error message is displayed.

- - - - - - - - - - -

2. Error message is displayed.

,

, ...

- - - - - - - - - - -

Further error messages are displayed, if present at the end of the list.

ERROR.LIST

ERRORLIST is displayed.

QUIT.NO

Do not delete error messages,
press

key three times = return to measuring mode.

QUIT.YES

Delete error messages.

ERROR QUIT

Error messages have been deleted.

ERROR.LIST

- - - - - - / - - -

Return to measuring mode.