KROHNE UFM 500 EN User Manual

Page 28

background image

28

5.8

Frequency output F

5.8.1

Application F (fct. 3.4.1)

Application F

Setting
via fct.3.4.1 or 3.3.1

Other functions
set via fct.3.4.6 to 3.4.8

Characteristic
of outputs,
see sect. 5.8.3

F
3.4.1

I
3.3.1

SMU F
3.4.6 to 3.4.8

1 flow direction

1. DIR

any

possible

F 1

F/R-operation
F/R-changeover via I

2 DIR

F/R IND.

possible

F 2

Direction indication for I

F/R IND.

2 DIR

possible

F 3

Switched off
(= 0 Hz / 0V)

OFF

any

no

F 4

Sound velocity
measurement

SOUND.VELO

any

no

F 5

For connection diagrams of outputs, see sect. 2.3.5.

5.8.2

Other functions for F

Fct. 3.4.2 Unit for frequency output F
PULSRATE

Setting pulses per unit of time (see fct. 3.4.3).

PULSE/UNIT

Setting pulses per unit of volume (see fct. 3.4.3).

Example of PULSRATE
Full -scale setting:

1,000 Liter per second (set via fct. 3.1.1).

Pulse rate:

1,000 pulses per second (set via fct. 3.4.3).

Pulse value:

1 pulse per Liter.

Changeover of full-scale setting:

2,000 Liter per second (change over via fct. 3.1.1).

Pulse rate:

unchanged (see above), 1,000 pulses per second.

Pulse value now:

1 pulse per 2 liter.

Example of PULSE/UNIT
Full-scale setting:

1,000 Liter per second (set via fct, 3,1.1).

Pulse value:

1 pulse per Liter (set via fct. 3.4.3).

at 1000 Liter per second:

1,000 pulses per second

± 1 pulse per Liter.

Changeover of full-scale setting:

2,000 Liter per second (change over via fct. 3.1.1).

Pulse value:

unchanged (see above) 1 pulse per Liter.

at 2000 Liter per second:

2,000 pulses per second

± 1 pulse per Liter as before.

Fct. 3.4.3 Pulse rate for 100% flowrate (F

100%

)

(appears only if “PULSRATE” set in fct. 3.2.2).
Range:

2.778

×

10

-3

to

1,000

PuLSe/Sec

0.1667

to

60,000

PuLSe/min

10

to

3,600,000

PuLSe/hr

Setting: Refer to sect. 5.3 “Special settings”!

Fct. 3.4.3 Pulse value
(appears only if “PULSE/UNIT” set in fct. 3.4.2).
Unit: Select from List in section 5.1.
Range: 0.0001 to 9.9999

×

10

9

PuLSe/Unit.

Setting: Refer to sect. 5.3 “Special settings”!
Setting is not checked but: Q

100%

×

pulse value must be less than or equal to 3,600,000 pulses/hr (equivalent to kHz)!

Fct. 3.4.5 Pulse width
Five pulse widths (30 / 50 / 100 / 200 / 500 m Sec) can be selected for programmed output frequencies (F

100%

, fct. 3.4.3)

less than or equal to 10 Hz (note output load and frequency ranges, see table in sect. 2.3.3).
Fixed pulse widths (50% duty cycle) are provided for programmed output frequencies above 10 Hz (see sect. 2.3.3)
regardless of the pulse width (see above) that has been set.

Fct. 3.4.5 Time constant for F
40 mSec

Time constant = 0.04 seconds (best for counting and/or batching processes).

SAME AS I

Same time constant as for current output I, see fct. 3.3.6.
(practical if frequency output F is used for instantaneous-value measurement).