6.2 Technical data

Measuring system

Measuring system

Measuring system

Measuring system

Measuring principle

Conductive conductivity

Measuring range

OPTISENS COND 1200 (W)

0.1...200 µS/cm (c=0.05)

1...2000 µS/cm (c=0.2)

1...20 mS/cm (c=1)

OPTISENS COND 1200 (PW)

0.05...10 µS/cm (c=0.01)

0.001...1 mS/cm (c=0.1)

OPTISENS COND 1200 (GF)

10 µS/cm...15 mS/cm (c=1)

Materials

Construction

OPTISENS COND 1200 (W):

Body: PVDF

Cell: Stainless steel (1.4571)

OPTISENS COND 1200 (PW):

Body: PVDF

Cell: Stainless steel (1.4571) or titanium

OPTISENS COND 1200 (GF):

Body: PVDF

Cell: Graphite (GF)

Sensor options

With integrated Pt100 temperature sensor

Process connection

OPTISENS COND 1200 (W):

G3/4 A male thread

G1/2 A male thread for immersion assembly

OPTISENS COND 1200 (PW/GF):

G3/4 A male thread

G1/2 A male thread

G1 A male thread

3/4-14 NPT male thread

Measuring accuracy

Conductivity accuracy

OPTISENS COND 1200 (W): +/- 2% full scale

1

OPTISENS COND 1200 (PW): +/-10% from the nominal value

1

OPTISENS COND 1200 (GF): +/-10% from the nominal value

1

Operating conditions

Temperature range

OPTISENS COND 1200 (W/PW):

0...+135°C / +32...+275°F

OPTISENS COND 1200 (GF):

0...+130°C / +32...+266°F

Max. operating pressure

OPTISENS COND 1200 (W/PW):

16 bar at +25°C, 9 bar at +60°C / 232 psi at +77°F, 130.5 psi at +140°F

OPTISENS COND 1200 (GF):

16 bar at +25°C, 9 bar at +60°C / 232 psi at +77°F, 130.5 psi at +140°F

1 Depending on the production conditions, the cell constant can deviate from the nominal value. This deviation can be

compensated at the signal converter.