Operation – KROHNE OPTISENS COND 1200 EN User Manual
Page 32
4
OPERATION
32
OPTISENS COND 1200
www.krohne.com
01/2014 - 4001967301 - MA OPTISENS COND 1200 R01 en
Step 3a: accessing the calibration menu via the main menu setup
setup
setup
setup
• You can start the calibration procedure now as described in "Step 4".
•
Press >>>> for more than 2.5 seconds, then release the button. You are on the main menu level. In the
upper line of the display "A" appears, beneath the main menu quick setup
quick setup
quick setup
quick setup is highlighted.
•
Press
or
until the main menu setup
setup
setup
setup is highlighted.
MAIN MENU
A quick setup
B test
> C setup
> C setup
> C setup
> C setup
D service
•
Press >
>>> to enter the chosen menu.
You are on the first submenu level. In the upper line of the display "setup" and "c1"
appears, beneath the submenu process input A
process input A
process input A
process input A is highlighted.
•
Press
or
to select process input A
process input A
process input A
process input A is highlighted.
•
Press >>>> to enter the chosen menu.
You are on the second submenu level. In the upper line of the
display "process input A" and "C1.1" appears, beneath the
submenu parameter
parameter
parameter
parameter is highlighted.
•
Press
or
until the submenu cell calibration
cell calibration
cell calibration
cell calibration is highlighted.
•
Press >>>> to enter the chosen menu.
.book Page 32 Friday, January 24, 2014 12:32 PM