Step 3a: accessing the calibration menu via the main menu setup

setup

setup

setup

• You can start the calibration procedure now as described in "Step 4".

Press >>>> for more than 2.5 seconds, then release the button. You are on the main menu level. In the
upper line of the display "A" appears, beneath the main menu quick setup

quick setup

quick setup

quick setup is highlighted.

Press

or

until the main menu setup

setup

setup

setup is highlighted.

MAIN MENU

A quick setup

B test

> C setup

> C setup

> C setup

> C setup

D service

Press >

>>> to enter the chosen menu.

You are on the first submenu level. In the upper line of the display "setup" and "c1"

appears, beneath the submenu process input A

process input A

process input A

process input A is highlighted.

Press

or

to select process input A

process input A

process input A

process input A is highlighted.

Press >>>> to enter the chosen menu.

You are on the second submenu level. In the upper line of the

display "process input A" and "C1.1" appears, beneath the

submenu parameter

parameter

parameter

parameter is highlighted.

Press

or

until the submenu cell calibration

cell calibration

cell calibration

cell calibration is highlighted.

Press >>>> to enter the chosen menu.